Pack 1233.zip (2024)
: It can identify air leaks, inadequate vacuum levels, or complete air removal failures within the system.
: The pack meets ANSI/AAMI/ISO 11140-5:2007 Class 2 standards for Bowie-Dick test packs. Pack 1233.zip
The test pack contains a chemical indicator sheet. A across the entire sheet indicates a "pass" (proper air removal), while any non-uniformity or spotty color change signifies a "fail," indicating that the sterilizer requires maintenance. 3M™ Comply™ Bowie-Dick Test Pack 1233LF (AAMI Class 2) : It can identify air leaks, inadequate vacuum
: It is typically used daily to monitor the performance of the sterilizer before the first processed load. Interpretation of Results A across the entire sheet indicates a "pass"
The 3M™ Comply™ Bowie-Dick Test Pack 1233LF is an equipment monitor designed to test the efficiency of air removal in steam sterilizers operating at . It is a single-use, disposable test pack that helps ensure the sterility of medical instruments by detecting potential mechanical failures. Key Functions