Often caused by the internal relay L1 . This relay frequently oxidizes, preventing a clean contact for short/open lead zeroing.

Some technicians successfully use DeoxIT spray, though opening the relay to clean contacts with a sliver of a business card soaked in alcohol is more effective. Repair Of Sencore Lc103 Service Manual

Aging electrolytic capacitors in the power supply section are a common culprit for random hangs or "frozen" screens. Often caused by the internal relay L1

If your leakage tests are inconsistent, check for corroded connectors between internal boards. Cleaning the board interconnects with contact cleaner often stabilizes the measurement circuit. Aging electrolytic capacitors in the power supply section

Perform a preventative "re-cap" of the PSU section if the unit is more than 20 years old. 📘 Essential Documentation Sources

Specific step-by-step PDF guides for calibrating the LC103 are hosted on the EEVblog Electronics Community Forum . ⚠️ Critical Safety Warning Sencore TC 162 wiring diagram needed for repair - Facebook